COMPACT HIGH-VOLTAGE POWER SUPPLIES
- Max. Voltage: ±1 to ±2 kV
- Max. Current: 100 to 200 mA
- Max. Power: 200 W
- Reversible output polarity
Easy polarity switching via front panel
The EJ series features reversible polarity that can be controlled using a knob on the front panel. This allows users to switch output polarity safely and easily without the need for rewiring, significantly improving efficiency during testing applications. Please contact us for information regarding high-power models.
FEATURES AND BENEFITS
- Reversible output polarity
- Front-panel polarity switching knob
- Safe polarity reversal without rewiring
- Dual current range for precise micro-current control (Optional)
- Ideal for withstand voltage testing of diodes and power semiconductors
- Compact benchtop design
APPLICATIONS
- Insulation breakdown test
- Electrostatic Discharge (ESD) Testing
- All kinds of High-voltage Testing
Models
| Model | Output | |
|---|---|---|
| Voltage | Current | |
| EJ-1R200 | 0 to +1/0 to -1 kV | 0 to 200 mA |
| EJ-1.5R133 | 0 to +1.5/0 to -1.5 kV | 0 to 133 mA |
| EJ-2R100 | 0 to +2/0 to -2 kV | 0 to 100 mA |
Specifications
- Input voltage
- 100 to 240 Vac, 50/60 Hz, Single-phase
- Output voltage control
-
[Local] 10-turn potentiometer on front panel
[Analog remote] External control voltage 0 to 10 Vdc - Output current control
-
[Local] 10-turn potentiometer on front panel
[Analog remote] External control voltage 0 to 10 Vdc
Options
- -LCr
-
Dual Current Range
Switchable between 100% and 10% of Full Scale (F.S.).
Switching the range changes both the meter display and the external control signal scaling (0 to 10 V) to improve resolution for low-current applications.
Example: 200 mA F.S. switches to 20.0 mA F.S.
- -LN
-
Power Failure Recovery
Automatically restores the high-voltage output state when AC power is restored effectively resetting the protection circuit.
- -LZ
-
Carrying Handle
A handle is attached to the top of the unit for portability.
How to Order
When ordering with options, please add the option names in alphabetical order following the model name, using the example below as a reference.
Example: EJ-2R100-LCrNZ
Accessories
| Standard | CABLE TYPE 1 | 125 V / 10 A | 2.5 meters Fixed length |
||
|---|---|---|---|---|---|
| Sold separately [for 200 V input] |
CABLE TYPE 3 | 250 V / 10 A | 2.5 meters Fixed length |
| Standard | CN-BNC-HVP | ![]() |
2.5 meters (Specified length available) |
|---|
Dimensions
Download
If you are unable to download a file
Please try the following solution.
- Please press Ctrl+F5 to clear the cache of your web browser and try again.
- Please restart your web browser and log in again to try again.
- Please change your web browser to another browser and try again.
- Restart the computer and try again.
- Please try again on a different computer.

-
EJ series Datasheet
Date: 2024-01-16 rev.09
PDF (979 KB)

-
Safety and Usage of High voltage Power supply
Date: 2025-09-08 rev 04
PDF (602 KB)

-
High Performance High Voltage Power Supplies
Date: 2023-10-06 rev. 12
PDF (3,886 KB)

-
EJ series Instruction Manual
Date: 2023-10-24 rev 0.3
PDF (402 KB)
Login Required

-
EJ series Datasheet
Date: 2024-01-16 rev.09
PDF (979 KB)

-
Safety and Usage of High voltage Power supply
Date: 2025-09-08 rev 04
PDF (602 KB)

-
High Performance High Voltage Power Supplies
Date: 2023-10-06 rev. 12
PDF (3,886 KB)

-
EJ series Instruction Manual
Date: 2023-10-24 rev 0.3
PDF (402 KB)
On this website, we provide only the latest versions of information and instruction manuals for our products. Therefore, the newest versions of manuals on the website may differ from those that came with products you purchased in the past.