Ring artifacts appear as concentric rings of varying brightness in CT images. These are primarily caused by inconsistent sensitivity across detector elements or defective pixels. To minimize these artifacts, perform detector calibration (gain correction) and utilize the system's built-in ring artifact correction functions.