What is the positive or negative common of an analog remote control?

Category
Product Support FAQ

The common terminal for the analog remote control is internally connected to either the positive or negative output terminal (sense/load), or to the product's enclosure (chassis ground). When the common is connected to the positive output terminal, it is referred to as "positive common." Conversely, when it is connected to the negative output terminal, it is called "negative common." Connecting the common terminals of a positive common product and a negative common product together will create a short circuit across their outputs. This will cause severe damage to the products and lead to failure. To prevent such issues, some products feature an analog control interface that is galvanically isolated from the power supply's output. This design provides enhanced safety and operational flexibility.
To avoid damaging your equipment, please consult the following table and the product's instruction manual to ensure you are using the correct configuration.

Products series Analog remote control Positive common Negative common Other common
R4K-36 series  -LRmf option Positive Sense
R3K-36 series -
R4K-80 series -
RK-80 series  -LRmf option Positive Sense
P4KF-80 series Negative Sense
TB series Negative Sense
PKTS series Galvanic Isolation
RK series Negative Sense
R4GT series Positive Sense
R4G series Positive Sense
R4GN series Negative Sense
RKT series Negative Sense
PRKT series Galvanic Isolation
PRT series Galvanic Isolation
PRTM series Galvanic Isolation
PVCE/PVCEJ series Negative Sense
REK/REKJ series Negative Sense
REKM series Negative Sense
RE series Negative Sense
REH series Negative Sense
REM series Galvanic Isolation
Rnet/RnetK/Rnet2K series Negative Sense
PBR series Galvanic Isolation
PBRM series Galvanic Isolation
PBH series Galvanic Isolation
EP series -
HS series Enclosure
EQ series Enclosure
EPR series Enclosure
EJ series Enclosure
CZE series Enclosure
HARS series Galvanic Isolation
AU series Enclosure
(except for -LF option)
AE series Enclosure
AK series Enclosure
(except for -LF option)
AKP series Enclosure
(except for -LF option)
AUH series Enclosure
(except for -LF option)
ASX series  -LA option Enclosure
AESS series Enclosure
HEB series -
HIB series -
XPg series Enclosure
XKg series Enclosure
SK series -
SKS series -
HECD series Enclosure
EJC series Enclosure
ECU series Enclosure
EC series Enclosure
ECA series Enclosure
EW series Negative Load Terminal
EGS series -
EGD series Negative Load Terminal
EZD series Negative Load Terminal