What is X-Ray Fluorescence? (Basic Knowledge)

What is X-Ray Fluorescence?

When an object is irradiated with X-rays of a certain energy or greater, the inner shell electrons of the atoms in the material are excited. When the outer shell electrons move to the holes that occur as a result of this, X-rays equal to the difference in energy between the inner and outer shells are released. This is called X-ray fluorescence.
The difference in energy between the inner and outer shells is unique for each element. Therefore, the energy of X-ray fluorescence is unique for each element. In addition, the energy is greater for shells that are closer to the nucleus. So, K shell > L shell > M shell. X-ray fluorescence that is emitted during a transition from the L shell to the K shell are called Kα rays, and the X-ray fluorescence that is emitted during a transition from the M shell to the K shell are called Kβ rays.

Principle of Measurement

Since the energy of X-ray fluorescence is unique for each element, we can examine the fluorescence spectrum of a sample to analyze its constituent elements. (Figure 2) Also, by measuring the intensity of the spectrum, we can also determine the concentration of specific elements in a sample.

Figure 2 Fluorescent X-ray spectrum (Sample)

Analysis Equipment

X-ray fluorescence analyzers can be broadly divided into two categories. One type are wavelength dispersive X-ray fluorescence analyzers (WDX) , which prepare X-ray fluorescence with Soller slits (extracting nearly parallel fluorescence emissions) and use spectroscopic crystals to take spectroscopic measurements. They have high detection sensitivity and high energy resolution, which makes them suitable for cases where a greater level of detection accuracy is required. The other type are energy dispersive X-ray fluorescence analyzers (EDX), which are usually equipped with a semiconductor detector. Since the semiconductor detector itself has a function to analyze energy, the features of this type include compact size, short analysis time, and simultaneous multi-element analysis.

Quantitative Analysis

Quantitative analysis is a method of determining the content of elements in a sample. For X-ray fluorescence analysis, this method is broadly divided into two categories: the calibration curve method and the fundamental parameter (FP) method.

Calibration curve method: This is a general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of standard samples of known concentration.

FP method: This method is used to devise a theoretical estimate of the composition of a sample based on the intensity of X-ray fluorescence from an unknown sample that is measured, based on the principle of the generation of X-ray fluorescence. The FP method is used to measure samples that do not have a calibration curve or whose composition is unknown.

Sample Preparation Method

Although X-ray fluorescence analysis allows a variety of samples, such as solids, liquids, and powders, to be analyzed non-destructively, there is a possibility that the analysis results may differ depending on the shape of the sample or how the sample is placed in the container. (Figure 3) Therefore, the following points are important when preparing samples.

Points for preparing samples

  • Prepare homogeneous samples
  • Prepare the standard sample and each sample to be measured in the same way
  • Avoid contamination
There is an effect of sample surface on generated fluorescent X-ray.
(a) Bulk sample
Rough: Sample surface are different on the primary X-ray side and fluorescent X-ray side.(b) Powder sample

Figure 3 Effects of specimen surface and granularity (particle size)

Regulations for Various Hazardous Substances

Attention is paid to measures against hazardous chemical substances as an environmental issue. RoHS directives and other regulations are emerging in order to prevent the diffusion of hazardous chemical substances. With such trends, X-ray fluorescence analyzers (XRF) that can perform screening analyses quickly and easily are required.

* Revisions may be made to the legal interpretations of these regulations. The statements here do not represent the positions of official institutions. Be sure to refer to the original text of applicable laws and regulations.

RoHS Directive

The RoHS (Restriction of Hazardous Substances) was issued by the EU in 2003, and came into force in July 2006.
Therefore, in order to ship products to EU member countries, the content of six types of hazardous substances specified in RoHS must be less than the specified values. (Table 1) X-ray fluorescence analyzers (XRF) are non-destructive and can perform measurements very quickly, which makes them useful for screening hazardous substances. In addition, XRF can be used to perform both qualitative and quantitative analyses. Hexavalent chromium is detected as total chromium, and PBB/PBDE are analyzed as total bromine.

table1 RoHS regulated 6 substance
Material Regulation value
Lead (Pb) 1000ppm less
Mercury (Hg) 1000ppm less
Cadmium (Cd) 100ppm less
Hexavalent chromium (Cr6+) 1000ppm less
Polybrominated biphenyl (PBB) 1000ppm less
Polybrominated biphenyl tail (PBDE) 1000ppm less

Soil Contamination Countermeasures Act

The Soil Contamination Countermeasures Act came into effect in February 2003 in order to prevent damage to health due to soil contamination and to gain a better understanding of contamination conditions. Among the specific hazardous substances that are specified in this law, the components of Class 2 heavy metals can be analyzed with X-ray fluorescence.
The specified substances and their reference values are shown in Table 2. An official quantification method is enacted only for total arsenic and total lead (JIS K 0470).

table2 Soil Contamination Countermeasures Act
Material Limit standard
Lead (Pb) 150mg/kg
Cadmium (Cd) 150mg/kg
Hexavalent chromium (Cr6+) 250mg/kg
Arsenic (As) 150mg/kg
Mercury (T-Hg) 15mg/kg
Selenium (Se) 150mg/kg

Regulations Regarding Toys

Toy safety is regulated by Part 3 of the EU Toy Safety Directive (EN71 Part 3).
It stipulates regulations regarding the oral intake of heavy metals in toys that are designed for children aged 6 or younger, and that could be placed in the mouth or swallowed. The target metal elements and their regulation values are shown in Table 3.
In Japan, Toy Safety Standards (ST-2002) have been established, and the regulation value for lead (90 ppm) was added in May 2006.
These ST standards include the Food Sanitation Act, EN71, and other regulations as inspection items.

Table3 Regulations Regarding Toys
Material Regulation value
Excluding clay and
finger point
Clay and finger point
Lead (Pb) 90ppm 90ppm
Mercury (Hg) 60ppm 25ppm
Cadmium (Cd) 75ppm 50ppm
Hexavalent chromium (Cr6+) 60ppm 25ppm
Barium (Ba) 1000ppm 250ppm
Arsenic (As) 25ppm 25ppm
Selenium (Se) 500ppm 500ppm
Antimony (Sb) 60ppm 60ppm

Halogen-Free Measures

Halogen compounds have been used widely in electronic materials and other products as fire retardant materials. However, because they generate harmful gas when burned (dioxin in particular can be produced at high temperature) and for environmental concerns such as preventing the destruction of the ozone layer, there is now a strong need to switch to halogen-free materials. However, there is no clear definition of "halogen-free".

For example, the Japan Electronics Packaging and Circuits Association (JPCA) has established regulation values for halogen content in printed wiring as a JPCA standard (Test Method for Halogen-Free Materials JPCA-ES01). It defines the total amount of chlorine (Cl) and bromine (Br) contained in epoxy resin and phenolic resin used in printed wiring boards. (Table 4)

This standard has also been adopted for IEC standard 61249-2-21 (International Electrotechnical Commission) and IPC standard 4101B (Association Connecting Electronics Industries).

table4 Total amount of chlorine and bromine in the JPCA standard

Material Limit standard
Chlorine (Cl) Total amount of 900ppm or less
Bromine (Br)

Relevant technical knowledge

Recommended product

News

News & Events
Technical Knowledge

Latest Articles

How to Use the Amplifier and Its Notes
What is X-Ray Fluorescence? (Basic Knowledge)
How to use DC Power Supply and remote sensing correctly and safely
The Right Way to Choose appropriate X-ray Inspection System
What is Clean Bench? (Basic Knowledge)
Product Search
Power Supplies
X-ray Inspection Systems
Analytical Instruments
News
News & Events
Technical Knowledge
Company
Company Details
Privacy Policy
Mission
History
Offices
Career
Career
Contact
How to order
Model selection assistance
Price information
Catalog request
Any question?
OEM custom design